ISO 80369 3 Annex A THE CONCERN Concerns have been raised about the possible risks of delivering accurate doses of medicines in certainclinical practices across high risk subpopulations e.g. neonatal patients when using a reversed connectionsystem female to male. This orientation may introduce inadvertent displacement of fluid
1Laboratory testing also shows a mid tolerance E
1 ENFitTM connector pair in a female to male orientationdisplaces a mean average of 0.148ml
1ISO 20695 Annexes A & K LOW DOSE TIP SYRINGE here called alternative enteralsyringe tip The ISO committee has decided to propose [it] for information only as test results lead to contradictoryconclusions on dose accuracy with this alternative enteral syringe tip
2 The design of this alternative enteral syringe tip may not ensure that the dose accuracy for enteral syringesis met; specific user information and education should accompany this alternative enteral syringe